Incubator
Product Configurations
Mode of Control
- Air temperature mode
- Skin temperature mode
- Humidity display and control
Temperature Control
- Air temperature control range: 25°C to 37°C (Override 37.1-39°C)
- Air temperature display range: 20°C to 45°C
- Skin temperature control range: 25°C to 37°C (Override 37.1-39°C)
- Skin temperature display range: 20°C to 45°C
- Accuracy of skin temperature sensor: 0.3°C
- Accuracy of air temperature sensor: ≤ 0.5°C
- Display resolution: 0.1°C
Humidity Control
- Humidity display range: 20%RH to 90%RH
- Humidity control range: 40%RH to 80%RH
- Humidity display precision: ±5%RH
- Humidity control precision: ±10%RH
Bed and Bassinet
- Bed surface temperature uniformity: ≤ 0.6°C
- Inclination of bassinet: +/- 12°
- Bassinet size: 630 mm x 340 mm
- Bassinet max load bearing: 100 N
- IV pole max load bearing: 15 N
- Tray rack max load bearing: 20 N
Additional Specifications
- Water tank volume: Over 1 L
- Noise inside hood: ≤ 60 dB (A)
- Power rate: 220 VAC, 50 Hz
- Input power: ≤ 550 VA
Failure Alarms
- Over Temperature Alarm
- Deviation Alarm
- Sensor Failure Alarm
- Power Failure Alarm
- System Failure Alarm
- Water Shortage Alarm
Additional Features
- Two round side windows and four operation windows
- Front hood door and extra back door
- Removable drawer-type water tank with ultrasonic nebulization for humidification
- RS-232 interface for data communication