NOVAMEDTEK with C-Arm

NOVAMEDTEK with C-Arm

TECHNICAL DATA

ESWL UNIT

  • Shockwave Source: Electromagnetic system
  • Coupling Type: Dry patient coupling
  • Shockwave Head Opening: 155 mm
  • Aperture Angle: 45°- 50°
  • Penetration Depth: 155 mm
  • Focal Pressure: 550 to 900 bar
  • Focal Area: ±20 mm x ±30 mm x ±25 mm

TREATMENT TABLE

  • All Movements: Motorized
  • Table Surface: 235 cm x 65 cm (with extension)
  • Up-Down Movement: 300 mm
  • Longitudinal Movement: 150 mm
  • Transverse Movement: 150 mm
  • Max Patient Load Capacity: 220 kg

NOVAREX NT-20 C-ARM IMAGING SYSTEM (OPTIONAL)

  • Image Intensifier: 9” (23 cm)
  • CCD Camera: 1K² technology
  • X-ray Power Output: 3.5 or 5 KW (Stationary or Rotating)
  • KV Range: 40 – 110
  • mA Range: From 0.2 up to 8.0

CONTROL UNIT

  • Type: High resolution medical grade
  • Monitor Sizes: 17” – 19” – 21”
  • Image Processing Software: Advanced computer system for imaging purposes
    • Features:
      • DVD/CD writer
      • DICOM interface (optional)
      • Patient database
      • Last Image Hold (LIH)
      • Automatic Brightness Control (ABC)

MOVABLE MONITOR TROLLEY

  • Power Voltage: 220-230 VAC, 50/60 Hz, 3.5 kVA, single phase
  • Grounding: 100Ω or less in grounding resistance
  • Operating Temperature: 10° C to 40° C
  • Safety Standard: IEC 601.1 class IIB
  • Space Requirement: 4m x 4m x 2.5m